What Is CT Resolution?

Resolution in computed tomography is not a single number, and the reason matters when choosing between techniques.

A reconstruction is built from voxels, which are the three-dimensional equivalent of pixels. Each voxel has dimensions in x, y, and z, where x and y describe the plane of the sample and z describes depth through it. Whether those three dimensions are equal turns out to be the defining difference between the two CT geometries used in electronics inspection.

Two geometries

In cone beam CT, projections are collected all the way around the sample. Medical scanners rotate the source and detector around a stationary patient. Industrial systems typically do the reverse, holding the source and detector fixed while the sample rotates a full 360 degrees.

Planar CT, also called laminography, works within a restricted angular range. The sample stays stationary while the source and detector orbit, collecting a set of 2D images that are reconstructed into slices.

Cone beam CT is isotropic

Because cone beam CT collects projections through a complete angular range, resolution is equal in all three directions. The voxel is a cube.

That means a feature can be measured just as precisely through the depth of the sample as across it, and the reconstruction can be sliced in any orientation with equal fidelity.

Planar CT is not

Planar CT produces excellent resolution in x and y, comparable to cone beam CT in the plane of the sample. Resolution in z is inferior.

The voxel is therefore not a cube but a rectangular prism, square in the sample plane and elongated through depth. How elongated depends on how many images are collected and across what angular range.

Why depth resolution depends on angular range

The reason is worth understanding, because it explains the entire tradeoff rather than just describing it.

Depth information in any tomographic reconstruction comes from parallax. Features at different heights shift by different amounts as the viewing angle changes, and that differential shift is what allows an algorithm to determine where along z a feature actually sits. The wider the range of angles collected, the more depth discrimination is available.

A limited angular range provides limited parallax, which provides limited depth discrimination. That is not an implementation shortcoming, it is a direct consequence of the geometry, and it is why z resolution in planar CT is a function of angular range and image count rather than a fixed specification of the machine.

What the tradeoff buys

Losing depth resolution sounds like a straightforward disadvantage. In practice, for flat samples, it frequently produces a better result.

The reason is magnification. Cone beam CT requires the sample to rotate fully between the source and the detector, which means it must be positioned far enough away to clear both throughout the rotation. On a large flat object like a populated circuit board, that clearance requirement forces the sample away from the source, and distance from the source is what limits magnification.

Planar CT has no such requirement. The sample stays still, so it can be brought very close to the source while the source and detector orbit around it. That produces very high magnification, and on a flat object the resulting in-plane image quality can exceed what cone beam CT achieves on the same sample.

Speed is the second gain, and it is dramatic. A planar CT BGA inspection runs in a couple of seconds. The same inspection by cone beam CT takes minutes, which is the difference between something that fits in a production cycle time and something that does not.

Choosing between them

Neither geometry is better in the abstract. They resolve different things well.

Cone beam CT is the right tool when depth measurement matters and the sample can rotate freely, which generally means smaller objects examined in a development or failure analysis context where time is available.

Planar CT is the right tool for flat, laterally large samples, where in-plane resolution and speed are the binding requirements and where full rotation is impossible anyway. That describes most electronics inspection, which is why it became the standard for BGA work.

The question is not which produces a better voxel. It is which dimension the inspection actually needs to resolve.

Creative Electron X-Ray Systems

Explore Industrial X-Ray Inspection & CT Systems

Compare Creative Electron X-ray inspection systems for PCB and SMT inspection, medical devices, industrial quality control, 2D X-ray, planar CT, 3D CT, automated inspection, AI analysis, and X-ray component counting.

Want to learn more?

Contact the X-ray People!

Contact Us
Intended Use

All information is confidential.

X-Ray University

Learn more from The X-Ray People!

Search our knowledge base or browse topics from the engineers and X-ray experts at Creative Electron.

Search X-Ray University